C201H Oxygen Transmission Rate Test System is developed based on the Coulometric oxygen analysis sensor and equal pressure method test principle. It is designed and manufactured according to ASTM D3985 and other standards to provide high precision and high efficiency oxygen transmission rate tests for high and medium gas barrier materials. It is suitable for oxygen permeability performance tests of film, containers, sheet and related materials in pharmaceutical, food, medical devices, daily chemistry, photovoltaic, electronics and other fields.
Prepared sample is clamped between the test chambers. Oxygen or air flows on one side of the film, and high pure nitrogen flows on the other side of the film. Oxygen molecule diffuses through the film to the other side with high purity nitrogen and is carried by the flowing nitrogen to the sensor. By analyzing the oxygen concentration measured by the sensor, oxygen transmission rate and other results can be calculated.
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003
This test instrument is applicable to the determination of oxygen permeability of:
Basic Applications |
---|
Oxygen transmission rate tests of various film, paper-plastic composite film, co-extrusion film, aluminum-plated film, aluminum foil composite film and glass fiber aluminum foil composite film,.etc. |
Oxygen transmission rate tests of bottles, pouches, cans, boxes and barrels of plastic, rubber, paper, paper-plastic composite, glass and metal, such as pharmaceutical packaging, wine bottles, coke bottles, Tetra Pak packaging, vacuum bags, three-piece cans, cosmetic packaging, toothpaste hose, jelly cups, yogurt cups. |
Oxygen transmission rate tests of PP, PVC, PVDC, metal foil, rubber, silicon, etc. |
Extended Applications |
Oxygen transmission performance tests of various packaging closures. |
Oxygen transmission performance tests of solar backpanel. |
Oxygen transmission performance tests of PPR tube and other materials. |
Overall oxygen transmission performance tests of medical blister systems. |
Oxygen transmission performance tests of plastic fuel tank. |
Oxygen transmission performance test of battery covers and separators. |
Parameter \ Model | C201H | |
---|---|---|
Test Range | cc/(m2·d) (Standard area is 50cm2) | 0.01 ~ 200 |
cc/(pkg·day) | 0.00005~1 | |
cc/(m2·d) (MASK area is 5cm2)) | 0.2~2000 (Optional) | |
Resolution | cc/(m2·day) | 0.0001 |
Repeatability | cc/(m2·day) | 0.01 or 1%, take the larger one |
Test Temperature | °C | 15~50 |
Temp. Fluctuation | °C | ±0.2 |
Test Humidity | RH | 0%, 35-90% ± 2% (Standard)100% (Optional) |
Additional Functions | GP-02 Gas Purification Device | Optional |
DataShieldTM Note3 | Optional | |
Computer System required by GMP | Optional | |
CFR21 Part11 | Optional |
Table 2: Technical Specifications
Test Chamber | 1 set for films + 1 set for containers |
Film Size | 4.2” x 4.2” (10.6cm×10.6cm) |
Film Thickness | ≤120 Mil (3mm) |
Container Size | ≤Φ 100 mm x 250mm; Φ15mm ≤ bottle mouth ≤ Φ 65mm |
Standard Test Area | 50 cm2 |
Carrier Gas | 99.999% nitrogen, 99.5% oxygen (Gas source is provided by the user) |
Carrier Gas Pressure | ≥ 40.6 PSI / 280 KPa |
Port Size | 1/8 inch metal tubing |
Dimensions | 26.7” H x 14.9” W x 22.8” D (68cm× 38cm× 58cm) |
Power | 120VAC ± 10% 60Hz / 220VAC ± 10% 50Hz (alternative) |
Net Weight | 150Lbs (68kg) |
Note 1: The described product characteristics are subject to the specific annotation of the "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions of the relevant laboratory environment standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
We can help you select or customize the proper unit for your test requirements.
Need help or have a question? For more information on how Labthink can help your company profit through testing, please Click the button.