C870M Integrated Loss-on-Drying Testing System is based on the gravimetric method and is designed and manufactured in accordance with pharmacopoeias and testing standards of the food and chemical industries. It is suitable for loss-on-drying testing in the above-mentioned fields.
Take the sample and mix thoroughly (if the crystals are large, they should be quickly crushed into particles smaller than 2 mm). Take approximately 1 g or the weight specified in the standard and place the sample in a test cup that has been dried to constant weight under the same conditions as the sample. Weigh accurately and, unless otherwise specified, dry at 105℃ to constant weight. Calculate the loss-on-drying of the test sample from the weight loss and the sample weight.
US Pharmacopeia, European Pharmacopoeia, British Pharmacopoeia, Japanese Pharmacopoeia, Chinese Pharmacopoeia, and other pharmaceutical-related testing standards
GB 5009.3-2016 and other food-related testing standards; GB/T 6284, JIS K 0068 and other chemical product-related testing standards.
| Basic Applications | Pharmaceuticals | Loss-on-drying and moisture content determination of various pharmaceuticals. |
| Extensive Applications | Food | Loss-on-drying and moisture content determination of various foods. |
| Chemicals | Determination of loss on drying for various chemical products. |
Table 1: Test Parameter Note 2
Parameter\Model | C870M | |
Test Range | mg | 0.3~40000 |
0.05~20000(Optional) | ||
Resolution | mg | 0.1 |
0.01 (Optional) | ||
Repeatability | mg | ±0.3 |
±0.05 (Optional) | ||
Temp. Range | ℃ | Room Temp. ~ 130 |
Temp. Fluctuation | ℃ | ±0.5 |
Test Pressure | kPa | 0 ~﹣20 Note 3 |
Extended Functions | 21 CFR Part 11 | Optional |
GMP Computer System Requirements | Optional | |
Table 2: Technical Specifications
Test Stations | 12 |
Test Cup Volume | 40 mLNote 4 |
Gas Specifications | Compressed air (air source is to be provided by the user) |
Gas Source Pressure | 0.5 MPa~0.7 MPa (72.5 PSI~101.5 PSI) |
Port Size | Φ8 mm polyurethane pipe |
Host Dimensions | 24.8” H x 41.3” W x 28.7” D (63cm× 105cm× 73cm) |
Power Supply | 120 VAC ±10% 60 Hz / 220 VAC ±10% 50 Hz (Select one from the two) |
Net weight | 396 Lbs (180 kg) |
Table 3: Product Configuration
Standard Configuration | Instrument host, balance (0.1 mg), liquid cooling module, gas drying module, test cups (12 pcs), Φ8 mm polyurethane tube |
Optional Parts | Software, GMP computer system requirements, 21 CFR Part 11, air compressor, test cup (40 mL), weight (200 g), balance (0.01 mg), weight (50 g) |
Note 1: All product features described are subject to the specific specifications listed in the"Technical Parameters" tables.
Note 2: All parameters in the table were measured in the Labthink laboratory by professional operators, according to the requirements and conditions of relevant laboratory environmental standards.
Note 3: Test pressure can be customized.
Note 4: Test cup volume can be customized, but the test range may vary and should be based on the actual delivery.
We can help you select or customize the proper unit for your test requirements.
Have a question or need assistance?
Discover how Labthink's testing instruments and packaging testing services can help your company improve efficiency and product quality. For expert support with packaging testing instruments, click the button to contact our experts and submit your inquiry today.