C401H Oxygen/Water Vapor Transmission Rate Test System is based on the testing principles of coulometric oxygen sensor and infrared water vapor sensor. It is designed and manufactured in accordance with standards such as ASTM D3985, ASTM F1249, and ISO 15106-2. It provides ahigh-precision integrated testing solution for water vapor and oxygen transmission rates of high and medium barrier materials. It is widely applicable in pharmaceuticals, food, medical devices, daily chemicals, photovoltaics, electronics and other fields, accurately detecting the oxygen and water vapor transmission performances of films, packaging, sheets, and related materials.
Product Features Note 1
Combining Tests of OTR and WVTR in One
Testing Principle
The pre-treated sample is clamped between the test chambers. Oxygen or nitrogen with stable relative humidity flows on one side of the film, and high-purity nitrogen flows on the other side. Oxygen or water molecules diffuse through the film into the high-purity nitrogen on the other side and are carried to the sensor by the flowing nitrogen. By analyzing the oxygen or water vapor concentration measured by the sensor, the oxygen or water vapor transmission rate and other results are calculated.
Reference Standards Note 1
ASTM D3985, ASTMF1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications Note 1
Basic Applications | Films | Oxygen and water vapor transmission rate tests of various plastic films, plastic composite films, paper-plastic composite films, co-extruded films, metalized films, aluminum foil composite films, glass fiber aluminum foil paper composite films, and other film materials. |
Packaging | Oxygen and water vapor transmission rate tests of bottles, pouches, cans, boxes, and barrels made of plastic, rubber, paper, paper-plastic composite, glass, and metal, including pharmaceutical packaging, wine bottles, cola bottles, peanut oil barrels, Tetra Pak packaging, vacuum packaging pouches, three-piece cans, cosmetic packaging, toothpaste tubes, jelly cups, and yogurt cups, etc. | |
Sheets | Oxygen and water vapor transmission rate tests for sheet materials such as PP sheets, PVC sheets, PVDC sheets, metal foil sheets, rubber sheets, and silicon wafers. | |
Extended Applications | Packaging Caps | Oxygen and water vapor permeability tests of the caps of various packaging. |
LCD Displays | Oxygen and water vapor permeability tests of liquid crystal displays and related films. | |
Solar Backsheet | Oxygen and water vapor permeability tests for solar backsheets and related materials. | |
Pipes | Oxygen and water vapor permeability tests for PPR pipes and pipes made of various materials. | |
Medical Blisters | Overall oxygen and water vapor permeability tests of medical blister packs. | |
Sterile Wound Dressings, Medical Plasters | Oxygen and water vapor permeability tests for sterile wound dressings, medical plasters, etc. | |
Battery Housings | Oxygen and water vapor permeability tests for battery housings |
Technical Parameters
Table 1: Test Parameters Note 2
Parameters\Model | C401H | |
Test Range | cc/(m2·day) (Standard area of 50 cm2) | 0.01~200 |
g/(m2·day) (Standard area of 50 cm2) | 0.005~50 | |
cc/(pkg·day) (Packaging) | 0.00005~1 | |
g/(pkg·day) (Packaging) | 0.00002~0.25 | |
cc/(m2·day) (MASK area of 5 cm2) | 0.2~2000 (Optional) | |
g/(m2·day) (MASK area of 5 cm2) | 0.2~500 (Optional) | |
Resolution | cc/(m2·day) | 0.0001 |
g/(m2·day) | 0.0001 | |
Repeatability | cc/(m2·day) | 0.01 or 1%, whichever is greater |
g/(m2·day) | 0.005 or 2%, whichever is greater | |
Temp. Range | ℃ | 15~50 |
Temp. Fluctuation | ℃ | ±0.2 |
Humidity Range | %RH | 0%, 35%~90%±2% |
Extended Functions | GP-02 Gas Purification Device | Optional |
DataShieldTM Note 3 | Optional | |
GMP Computer System Requirements | Optional | |
CFR 21 Part 11 | Optional | |
Table 2: Technical Specifications
Test Cells | One set for film and one set for container |
Film Size | 4.2” x 4.2” (10.6 cm × 10.6 cm) |
Sample Thickness | ≤120 Mil (3 mm) |
Container Dimensions | ≤Φ100 mm × 250 mm; Φ15 mm ≤ bottle mouth ≤ Φ65 mm |
Test Area | 50 cm2 |
Carrier Gas Specification | 99.999% high-purity nitrogen (gas source is to be provided by the user) |
Gas Source Pressure | ≥ 40.6 PSI / 280 KPa |
Port Size | 1/8” metal pipe |
Dimensions | 26.7” H x 15.7” W x 26.7” D (68cm × 40cm × 68cm) |
Power Supply | 120VAC ± 10% 60Hz / 220VAC ± 10% 50Hz (select one from the two) |
Net Weight | 165 Lbs (75 kg) |
Table 3: Product Configuration
Standard Configuration | Instrument host, sampler, vacuum grease, Φ4 mm polyurethane tube |
Optional Parts | GP-02, software, computer, GMP computer system requirements, CFR 21 Part 11, air compressor, DataShieldTM Note 3 |
Note | The compressed air inlet of this instrument is a Φ4 mm polyurethane pipe (pressure ≥ 79.7 PSI / 550 KPa); the air source is to be provided by the user. |
Note1: All product specifications shall be subject to the exact markings in the"Technical Parameters" table.
Note2: All parameters in the table were measured by professional operators at Labthink Laboratory in compliance with relevant laboratory environmental standards and conditions.
Note3: The DataShieldTM System provides secure and reliable data application support. This system can be shared by multiple Labthink products. Please purchase optionally as needed.
Have a question or need assistance?
Discover how Labthink's testing instruments and packaging testing services can help your company improve efficiency and product quality. For expert support with packaging testing instruments, click the button to contact our experts and submit your inquiry today.