C910H Integrated Air Permeability, Puncture and Thickness Testing System employs the volumetric method for gas permeability testing and mechanical principles for puncture strength and thickness measurements. It is specifically suitable for the measurement of air permeability, puncture strength, and thickness of battery separators.
Fully Automatic High-Precision Air Volume Determination Method
Contact-type Thickness Measurement Method
Mechanical-type Puncture Strength Measurement Method
Integrated Measurement with Automatic Sample Feeding
Intelligent Control & Reliable Performance
Air permeability performance is tested in an environment with controlled temperature, humidity, and normal atmospheric pressure. The test measures the time taken for 100 mL of air to permeate through an area of6.45 cm2 of the separator sample under a pressure of 1.21 kPa applied by the instrument.
ISO 5636, TAPPI T460, GB/T 36363-2018, GB/T 458-2008,GB/T 23227, GB/T 12655
Basic Applications | Air Permeability | Suitable for air permeability testing of battery separators, breathable films, and various other materials, as well as related polymer products. |
Thickness | Suitable for thickness measurement of battery separators, breathable films, and various other materials, as well as related polymer products. | |
Puncture Strength | Suitable for puncture strength measurement of battery separators, breathable films, and various other materials, as well as related polymer products. |
Table1: Test Parameters Notes 2
Parameters/Model | C910H | |
Measuring Range of Air Permeability | s/100 mL | 30~1500 |
um/(Pa · s) | 0.1~4 | |
Volume Resolution | mL | 0.001 |
Volume Precision | mL | 1% |
Pressure Range | kPa | 0~2 |
Pressure Precision | kPa | 0.01 |
Thickness Measuring Range | mm | 0~2 |
Resolution | um | 0.1 |
Repeatability | um | 0.5 |
Measuring Range of Puncture Strength | N | 0.1~50 |
Resolution | N | 0.001 |
Strength Value Precision | % | ±0.5 (Reading) |
Test Speed | mm/min | 0–500 (Any integer) |
Speed Accuracy | % | ±0.5 (Reading) |
Measuring Interval | mm | 0~600 |
Sample Feeding Speed | mm/s | 0~50 |
Extended Functions | GMP Computer System Requirements | Optional |
DataShieldTM Data Shield Note 3 | Optional | |
CFR 21 Part 11 | Optional | |
Table 2: Technical Specifications
Test Cell | 1 set |
Sample Size | 4” x 40” (10cm×100cm) |
Sample Thickness | ≤80 mil (2 mm) |
Measuring Area | 1 in2(6.4cm2) |
Thickness Measuring Head | 1 set |
Head Diameter | Φ3 mm (spherical) |
Measuring Head Pressure | <0.8N |
Puncture Fixture | 1 Set |
Fixture Inner Diameter | Φ50mm |
Puncture Needle Diameter | Φ1mm |
Air Source Specifications | Dry air (air source is provided by the user) |
Air Source Pressure | 72.5 PSI~101.5 PSI / 0.5 MPa~0.7 MPa |
Port Size | Φ6 mm polyurethane pipe |
Dimensions | 29.9” H x 17.7” W x 21.7” D (76cm × 45cm × 55cm) |
Power Supply | 120 VAC ± 10%, 60 Hz / 220 VAC ± 10%, 50 Hz (select one from the two) |
Net Weight | 187 lbs (85 kg) |
Table 3: Product Configuration
Standard Configuration | Main unit, embedded tablet computer, professional software, Φ6 mm polyurethane pipe |
Optionals | Air compressor, CFR 21 Part 11, GMP computer system requirements; DataShield? Data Shield Note 3 |
Remarks | The compressed air inlet of this unit uses a Φ6 mm polyurethane pipe (72.5–101.5 PSI/0.5–0.7 MPa). Air source is provided by the user. |
Note 1: The reference standards, test applications, and product specifications shall all follow the specific descriptions in the "Technical Parameters". "Reference Standards" refer to the testing method that is substantially consistent with the method described in the standards.
Note 2: All parameters in the table were measured in the Labthink laboratory by professional operators under the required conditions and in accordance with relevant laboratory environmental standards.
Note 3: The DataShield? Data Shield System provides secure and reliable data application support that can be shared across multiple Labthink products. Please purchase optionally based on your needs.
Note: Labthink is committed to continuous innovation and improvement of product performance and functions. For this reason, product technical specifications and appearance are subject to change without prior notice. The company reserves the right to make modifications, and the final right of interpretation.